Error Rates in an Indian Clinical Laboratory - Blog - Westgard QC

Tools, Technologies and Training for Healthcare Laboratories

Error Rates in an Indian Clinical Laboratory

Posted by Sten Westgard, MS

Earlier we discussed error rate issues at the Point-of-Care. But we didn't want to leave the "regular" laboratory out of the fun, so here's a study of error rates that came out in 2010:

Evaluation of errors in a clinical laboratory: a one-year experience, Goswami B, Singh B, Chawla R, Mallika V, CCLM 2010;48(1):63-66.

-----

Here is the description of the collection of data on the errors:

"From July 2008 to June 2009, a total of 67,438 routine venous blood specimens were received in the laboratory. Errors were detected in 954 samples, with a total error rate of 1.4%."

Below is a table with the error rates, plus we've added a Sigma-metric conversion, using the short-term Sigma scale table.

Error type # defects % defects/
total tests
Sigma-metric
Overall Pre-analytical 736 1.1% 3.8
Hemolyzed sample 508 0.7% 4
Insufficient sample 72 0.17% 4.5
Lipemic sample 7 0.01% 5.3
Incorrect identification 44 0.06% 4.8
Empty tube 10 0.01% 5.2
Requisition slip
without sample
16 0.23% 5
Illegible hand writing 69 0.1% 4.6
Tube broken
in centrifuge
6 0.009% 5.3
Physician's request
order missed
4 0.006% 5.4
Total Analytical Error Rate* 75 0.11% 4.6
Non-conformity with QC 6 0.009% 5.3
Random error 15 0.02% 5.1
Calibration drift 10 0.014% 5.2
Reagent contamination 8 0.012% 5.2
Systematic error: probe
lamp, blocked tubing
36 0.05% 4.8
Total Postanalytical Error Rate 143 0.21% 4.4
Transcription errors 112 0.17% 4.5
Prlonged TAT 31 0.05%7% 4.9

[* We added an asterix here because, as with most error studies, we do not believe the analytical error rate was correctly measured. In broad studies of this nature, it is likely that the same control limits were used to for all tests, which might either be too wide or too narrow (because the rate is so low, we suspect too wide). The proper measurement of Sigma-metrics for the analytical performance would be to set a quality requirement and measure imprecision and bias, then calculate the Sigma-metric (long term Six Sigma scale).

Another issue about the analytical error rate measurement: a single error in calibration is a one-to-many error. That is, one calibration drift could affect entire runs, not just a single instance. Hemolyzed specimens are a situation where one error affects one specimen. Thus, the comparison of errors here may not accurately reflect the impact of the different types of errors.]

What's even more interesting is to compare these error rates with those of the recent April 2011 study of error rates tracked over 5 years for a group of Spanish laboratories.

There are some significant differences:

  • The median error rate of controls exceeding acceptance limits was 3.4% or 3.4 Sigma, vs. the Indian laboratory having a (combined) control limit violation rate of 0.11% or 4.4 Sigma [again, possibly evidence that the counting of control violations is not standardized across labs]
  • The median error rate of (internal) delayed TAT for the Spanish study was 0.5% or 4.1 Sigma, vs. the Indian laboratory having a TAT rate of 3.2% and a 4.9 Sigma.
  • The median error rate of insufficient samples was 0.3% or 4.3 Sigma, vs. the Indian laboratory with a rate of 0.17% or 4.5 Sigma
  • The (combined) median incorrect identification rate for the Spanish labs was 3.51% or 3.4 Sigma, vs the Indian laboratory error rate of 0.06% or 4.8 Sigma

There was one interesting similarity, however:

  • Hemolyzed specimen rate was similar for both institutions, about 0.6% or 4.1 Sigma for the Spanish labs, and 0.7% or 4.0 Sigma for the Indian laboratory

The result of these studies shows that error rates can vary widely. sometimes by a factor of 58. Not all of this variation in error rates is from the differences in measuring and definition of errors. There's probably a difference in some practices as well. While analytical errors exhibit a closer degree of uniformity (the box is designed to operate the same way in all labs), every laboratory has its own unique pre-analytical and post-analytical context.

Thus, it may not be useful to generalize the occurrence rates of pre-analytical errors, or declare that one particular error type is more frequent than other types.

The morale of the story: your error rates will vary.

 

×
Stay Informed

When you subscribe to the blog, we will send you an e-mail when there are new updates on the site so you wouldn't miss them.

Rejection (error) rates in a Thai laboratory
Error Rates at the POC
 

Comments

No comments made yet. Be the first to submit a comment
Already Registered? Login Here
Monday, 29 June 2026

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input