Rong Regulation - Westgard QC - Blog

Tools, Technologies and Training for Healthcare Laboratories

Q & A: The Rong Regulation, Turkey Mis-Totaling Total Error

Posted by Sten Westgard, MS

Often we get questions about the problems that laboratories encounter when they wrestle with their local regulations.

Here we have a case where there are Turkish laboratory regulations that are irrational. In the words of the submitter:

"I have a question about the estimation of total CV. We have a requlation about total allowable error in Turkey....It is Turkish, but I strongly believe that you can follow it.

"The marked page can be seen below: There is an example for glucose. Two levels of controls are measured and the CV%’s calculated. The total CV are calculated from the CV’s of two different levels as seen in the equation. I can not accept it, because the levels are different even if the CV’s are close."

TurkeyTotalErrorCombination

What's gone Rong here? More after the jump...

-----
Continue reading
  597 Hits

Losing ground on HbA1c

Posted by Sten Westgard, MS

Continue reading
  703 Hits

NY Times Editorial Focuses on Failures in Medical Device Regulation

A recent New York Times Sunday editorial focused on the deplorable state of medical device regulation. We simply can no longer trust FDA clearances, particularly 510ks:

Continue reading
  772 Hits

A solution to IVDR over-regulation?

 The fallout of the IDVR continues. An interesting quote from a recent paper: "Although the EC aimed at bringing safe and effective medical tests to the EU-market, ~9,000 tests out of ~40,000 tests got lost in transition." Nearly a quarter of tests suddenly became unacceptable? That places IVDR at just above 2.2 Sigma. The proposed soluti...

Continue reading
  500 Hits

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input