Questions and Answers - Westgard QC - Blog - Page 2 - Results from #20

Tools, Technologies and Training for Healthcare Laboratories

Q & A: The Rong Regulation, Turkey Mis-Totaling Total Error

Posted by Sten Westgard, MS

Often we get questions about the problems that laboratories encounter when they wrestle with their local regulations.

Here we have a case where there are Turkish laboratory regulations that are irrational. In the words of the submitter:

"I have a question about the estimation of total CV. We have a requlation about total allowable error in Turkey....It is Turkish, but I strongly believe that you can follow it.

"The marked page can be seen below: There is an example for glucose. Two levels of controls are measured and the CV%’s calculated. The total CV are calculated from the CV’s of two different levels as seen in the equation. I can not accept it, because the levels are different even if the CV’s are close."

TurkeyTotalErrorCombination

What's gone Rong here? More after the jump...

-----
Continue reading
  599 Hits

Q&A: 10 Questions on QC from the first webinar

Posted by Sten Westgard, MS

Continue reading
  1242 Hits

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input